An Explosion Prone Chemist's Guide to: HPLC
If one is studying chemistry and dreams about working in a laboratory wearing a white coat, whether in industry or academia, we have been plagued by the word: HPLC. Either in our theory, practical, or internship, we have learned, touched, or at least heard of HPLC. HPLC is a hot cake selling at a fair—efficient, fast, and optimizable. Now, I have seen so many articles on the principles and workings of each part of the HPLC, but when it comes to using this instrument in real life, finding a step-by-step guide is a nightmare. The change in software in different situations makes it more difficult to find the right method. That is why the best method of learning HPLC is to be able to use it hands-on every time. So, why am I writing this if the learning is situation-based? Because, even when the software, computer, model, and mobile phases change from machine to machine, the principles of the steps remain the same. This guide is something I came up with to help me during my internship, and I am positive that this will be helpful for someone like me who CANNOT keep listening to someone while they are giving instructions.
Wait, but what is HPLC? For our newbie chemists, HPLC (High-Performance Liquid Chromatography) is a chromatographic technique, but instead of using a column and touching around a bunch of liquids, we can feed the analyzing samples into the machine, and it gives out the data from the detector in the machine. HPLC is used for the separation, analysis, and detection of a compound either in a mixture or in a solution, similar to traditional liquid-liquid column chromatography. However, HPLC is more robust and sensitive compared to its traditional counterparts.

The components of HPLC include: mobile phase (the solvents), sampler, column, detector, and the computer (Fig 2). The mobile phase carries the sample through the column where the components interact and come out via the detector, where each of the components is detected as it comes out at various times (retention time). The amount of each component can be quantified by the area under the curve of the chromatogram (the graph where the peaks of the components are visualized).
So, how and where to start with all the parts going on in the machine? This is a simplified workflow from my rough lab notebook.
Step 1: Switch the machine on; it's time to start the day with a long sequence of samples. This is also the time to prepare your samples as per your SOP and prepare the mobile phase. There should be 4 solvents: in my case, two solvents were to clean my column, and the main solvents were aqueous (line A) and organic (line B). I know multiple steps are going on here, but once all these are done, we can proceed to the actual machine.
Step 2: Put the solvents in the lines. Cover any holes on the cap (if there are any) to prevent any volatile organic solvent from escaping (if any is used).
Step 3: Make sure the machine is not running! This is to prevent increased flow speed and high pressure from bursting in the column. This is done to reduce the damage to the column. All the control from now on will be done on the panels in the software. Switch on the sampler, oven, and detector (UV-visible spectroscope lamp in my case). The playground is now on the column panel. Any changes will be done by right-clicking on that panel.
Step 4: In the panel, the machine is run to clean the column. In the meantime, while the column is being cleaned, you can load your samples and set up the sequence. If you have not changed/set up your mobile phase in step 1 or 2, this is the time to do so. The flow rate should be at a minimum all this while.
Step 5: Once the column is cleaned, the next step is to purge the lines intended for the sample sequence. This step is crucial because, with the change of the mobile phase, air bubbles might become trapped in the lines during solvent changes. Purging removes these bubbles, preventing them from reaching the column and causing it to dry out. There is a valve at the bottom of the HPLC that remains open. Purging is performed at a higher flow rate to expel the solvent from the lines. Each line is individually placed in 100% of the solvent during this process.
Step 6: The filling in the bottles are input in the system so that the software can keep a track of the solvent in the bottles. After purging, the method is input in the panel with a low flow rate. The valve is then closed which will rapidly increase the pressure in the machine.
Step 7: After the pressure is stabilised, we can rump up the flow if our method demands for higher flow rate. After rumping up the flow rate, we need to wait till the pressure gets stabilised. After the pressure is stabilised we can run the sequence and experiment is done after the end of the sequence.
Note:
Don't forget to put blank sample at the beginning and at the end of the sequence.
Make sure the samples and the blanks are put in the sequence as the right place on the sampler.
The injector needle washer should be filled and kept at the proper place.
This is not the ultimate guide to HPLC. These are very personal steps according to my experiment. However, the workflow should be similar to all the HPLCs. The interface of the software of HPLC will be different place to place. I was working in an Agilent HPLC in their respective software.